EasyManua.ls Logo

Fluke 5500A - Equipment Required for Calibration and Verification

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC600 Option
Equipment Required for Calibration and Verification
6
6-15
6-27. Equipment Required for Calibration and Verification
Table 6-15 lists the equipment, recommended models, and minimum specifications
required for each calibration and verification procedure.
Table 6-15. SC600 Calibration and Verification Equipment
Wave Generator and Edge Amplitude Calibration, AC Voltage and TD Pulser Verification
Instrument Model Minimum Use Specifications
Digital
Multimeter
HP 3458A
Voltage
1.8 mV to ±130 V p-p Uncertainty: 0.06%
Edge 4.5 mV to 2.75 V p-p Uncertainty: 0.06%
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1% (used with Edge Amplitude
Calibration and AC Voltage Verification)
BNC Cable (supplied with SC600)
Edge Rise Time and Aberrations Verification
High-
Frequency
Digital Storage
Oscilloscope
Tektronix 11801 with
Tektronix SD-22/26
sampling head, or
Tektronix TDS 820 with
8 GHz bandwidth
Frequency 12.5 GHz
Resolution 4.5 mV to 2.75 V
Attenuator
Weinschel 9-10 (SMA)
or Weinschel 18W-10
or equivalent
10 dB, 3.5 mm (m/f)
Adapter
BNC(f) to 3.5 mm(m)
BNC Cable (supplied with SC600)
Leveled Sine Wave Amplitude Calibration and Verification
AC
Measurement
Fluke 5790A
Range 5 mV p-p to 5.5 V p-p
Standard
Frequency 50 kHz
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1%.
BNC Cable (supplied with SC600)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Digital
Multimeter
HP 3458A
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1%.
BNC Cable (supplied with SC600)

Table of Contents

Other manuals for Fluke 5500A

Related product manuals