EasyManua.ls Logo

Fluke 5500A - Page 142

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Service Manual
6-16
Table 6-15. SC600 Calibration and Verification Equipment (cont.)
Pulse Width Calibration and Verification
High-Frequency Digital
Storage Oscilloscope
Tektronix 11801 with Tektronix SD-
22/26 sampling head
Attenuator 3 dB, 3.5 mm (m/f)
Adapter (2) BNC(f) to 3.5 mm(m)
BNC Cable (supplied with SC600)
Leveled Sine Wave Frequency Verification
Frequency
Counter
PM 6680 with option (PM 9621, PM 9624,
or PM 9625) and (PM 9690 or PM 9691)
50 kHz to 600 MHz, <0.15 ppm
uncertainty
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC600)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
AC Measurement Fluke 5790A Range 5 mV p-p to 5.5 V p-p
Standard with -03 option Frequency 50 kHz to 10 MHz
Adapter
Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC600)
Leveled Sine Wave Harmonics Verification
Spectrum Analyzer HP 8590A
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC600)
Pulse Period, Edge Frequency, AC Voltage Frequency Verification
Frequency Counter PM 6680 with option (PM
9690 or PM 9691)
20 ms to 150 ns, 10 Hz to 10 MHz: <0.15 ppm
uncertainty
BNC Cable (supplied with SC600)
Edge Duty Cycle
Frequency Counter PM 6680
BNC Cable (supplied with SC600)
Overload Functional Verification
Termination Feedthrough 50 ± 1%.
BNC Cable (supplied with SC600)
MeasZ Resistance, Capacitance Verification
Resistors 1 M and 50 nominal values
Capacitors 50 pF nominal value at the end of BNC(f)
connector
Adapters to connect resistors and capacitors to BNC(f)
connector
BNC Cable (supplied with SC600)

Table of Contents