Calibration and Verification
Calibration
3
3-7
3-9. DC Current
Use a precision DMM and appropriate precision shunts to measure the 5500A output as
Figure 3-2 shows. Enter into the 5500A each of the measured values listed in Table 3-4
when prompted to do so.