6-66. Wave Generator Verification
This procedure uses the following equipment:
5790A AC Measurement Standard
BNC(f) to Double Banana adapter
50 feedthrough termination
BNC cable supplied with the Calibrator Mainframe
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
SC600
Cable
5500A-SC600
50
Feed Through
Termination
BNC (F) to
Double Banana
Adapter
om060f.eps
Figure 6-13. Wave Generator Verification Setup
For wave generation verification procedures, refer to Figure 6-13 for the proper setup
connections.
Set the Calibrator Mainframe to SCOPE mode, with the Wavegen menu on the display.
Press O on the Calibrator Mainframe to activate the output. Set the offset to 0 mV,
and the frequency to 1 kHz. Then follow these steps to verify the wave generator
function.
6-67. Verification at 1 M
Set the Calibrator Mainframe impedance to 1 M (The blue softkey under SCOPE Z
toggles the impedance between 50 and 1 M).
1. Connect the BNC cable to the Calibrator Mainframe’s SCOPE connector. Connect
the other end of the BNC cable to the 5790A INPUT 2 using the BNC(f) to Double
Banana adapter.
2. Set the 5790A to AUTORANGE, digital filter mode to FAST, restart fine, and Hi
Res on.
3. Program the Calibrator Mainframe to output the wave type and voltage listed in
Table 6-41.
4. Allow the 5790A reading to stabilize, then record the 5790A rms reading for each
wave type and voltage in Table 6-41.