5500A
Service Manual
6-98
Table 6-59. Leveled Sine Wave Frequency Verification
Calibrator Mainframe
Frequency
PM 6680 Settings PM 6680 Reading Tolerance
(output @ 5.5 V p-p) Channel Filter (Frequency)
50 kHz A On 1.25 Hz
500 kHz A Off 12.5 Hz
5 MHz A Off 125.0 Hz
50 MHz A Off 1250 Hz
500 MHz C Off 12500 Hz
6-126. Leveled Sine Wave Harmonics Verification
This procedure uses the following equipment:
Hewlett-Packard 8590A Spectrum Analyzer
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC300
Refer to Figure 6-24 for proper setup connections.
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
HP 8590
SC300
Cable
BNC(F)
to Type N (M)
Adapter
5500A-SC300
om066f.eps
Figure 6-24. Leveled Sine Wave Harmonics Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Levsine menu on the display.
Then follow these steps to verify the leveled sine wave harmonics.
1. Using the BNC cable and BNC(f) to Type N(m) adapter, connect the SCOPE
connector on the Calibrator Mainframe to the HP 8590A.
2. Program the Calibrator Mainframe to 5.5 V p-p at each frequency listed in Table 6-
60. Press O on the Calibrator Mainframe to activate the output.