EasyManua.ls Logo

Fluke 5500A - Remote Commands for 5500 A Calibration

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Service Manual
3-16
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
NORMAL AUX
V, ,
RTD
A, -SENSE,
AUX V
STBY
0.1 Ohm shunt placed as closely as possible
to the AUX terminals of the 5500A
If the Phase Meter LO terminals are not common
use a short between NORMAL LO and AUX LO
on the 5500A
Signal
Terminals
5500 A
Clark-Hess
Phase Meter
Reference
Terminals
HI
LO
TC
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
V, ,
RTD
A, -SENSE,
AUX V
om015f.eps
Figure 3-8. Volts and Current Phase Calibration
Table 3-11. Volts and Current Phase Calibration Steps
Reference Signal
Step NORMAL Output ( Volts) Current Output (Amps) Frequency (Hz)
(0° phase)
1 3.00E+00 300E-03 500E+000
2 3.00E+00 300E-03 10E+3
3 3.00E+00 2.00E+00 500E+00
4 3.00E+00 2.00E+00 5.0E+03
5 3.00E+00 3.00E+00 64E+00
6 3.00E+00 3.00E+00 1.0E+03
3-19. Remote Commands for 5500A Calibration
Calibration of the 5500A using remote commands is simple. To access calibration steps
described in paragraphs 3-6 through 3-15, simply send the command:
CAL_START MAIN
To access calibration steps described in paragraphs 3-16 through 3-18, send the
command:
CAL_START FACTORY