SC600 Option
Verification
6
6-59
5. Allow the Calibrator Mainframe reading to stabilize, then record the Calibrator
Mainframe capacitance reading for each nominal value listed in Table 6-46.
Compare the Calibrator Mainframe capacitance readings to the actual capacitance
values and the tolerance column of Table 6-46.
Table 6-46. MeasZ Capacitance Verification
Nominal
Capacitance Value
Calibrator
Mainframe
Capacitance
Reading
Actual
Capacitance Value Tolerance
5 pF 0.75 pF
29 pF 1.95 pF
49 pF 2.95 pF
6-73. Overload Function Verification
This procedure uses the following equipment:
• 50 Ω feedthrough termination
• BNC cable supplied with the Calibrator Mainframe
Refer to Figure 6-14 for setup connections.