Introduction and Specifications
Specifications
1
1-21
1-15. Power and Dual Output Limit Specifications
Frequency Voltages
(NORMAL)
Currents Voltages
(AUX)
Power
Factor
(PF)
DC 0 to ± 1020 V 0 to ± 11 A 0 to ± 3.3 V
10 to 45 Hz 33 mV to 32.9999 V 3.3 mA to 2.19999 A 10 mV to 3.3 V 0 to 1
45 to 65 Hz 33 mV to 1020 V 3.3 mA to 11 A 10 mV to 3.3 V 0 to 1
65 to 500 Hz 330 mV to 1020 V 33 mA to 2.19999 A 100 mV to 3.3 V 0 to 1
65 to 500 Hz 3.3 V to 1020 V 33 mA to 11 A 100 mV to 3.3 V 0 to 1
500 Hz to 1 kHz 330 mV to 1020 V 33 mA to 11 A 100 mV to 3.3 V 1
1 to 5 kHz 3.3 V to 1020 V [1] 33 mA to 2.19999 A 100 mV to 3.3 V [1] 1
5 to 10 kHz 3.3 V to 1020 V [2] 33 mA to 329.99 mA 1 V to 3.3 V [2] 1
[1] In dual volts, voltage is limited to 3.3 to 500 V in the NORMAL output.
[2] In dual volts, voltage is limited to 3.3 to 250 V in the NORMAL output.
• The range of voltages and currents shown in “DC Voltage Specifications,” DC Current Specifications,”
“AC Voltage (Sinewaves) Specifications,” and “AC Current (Sinewave) Specifications” are available in
the power and dual output modes (except minimum current for ac power is 0.33 mA). However, only
those limits shown in this table are specified. See “Calculating Power Uncertainty” to determine the
uncertainty at these points.
• The phase adjustment range for dual ac outputs is 0 to ± 179.99 degrees. The phase resolution for
dual ac outputs is 0.02 degree.
1-16. 5500A Phase Specifications
1-Year Absolute Uncertainty, tcal ± 5 °C, (
∆Φ
degrees)
10 to 65 Hz
0.15° [1]
65 to 500 Hz
0.9° [2]
500 to 1 kHz
2.0° [3]
1k to 5 kHz
6°
5k to 10 kHz
10°
[1] For 33 to 1000 V output, burden current <6 mA. For 6 to 20 mA burden current (33 to 330 V), the phase
uncertainty is 0.4°.
[2] For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase
uncertainty is 1.5°.
[3] For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase
uncertainty is 5°.