Introduction and Specifications
Additional Specifications
1
1-27
1-22. AC Voltage (Non-Sinewave) Specifications
Trianglewave &
Truncated Sine
Ranges
Frequency
1-Year Absolute Uncertainty,
tcal ± 5 °C,
± (% of output + % of range) [2]
Maximum
Voltage
Resolution
p-p [1]
%Output %Range
Normal Channel (Single Output Mode)
2.9 to 92.999 mV 0.01 to 10 Hz 5.0% 0.5% Two digits on each range
93 to 929.999 mV 10 to 45 Hz 0.25 0.5
0.93 to 9.29999 V 45 Hz to 1 kHz 0.25 0.25 Six digits on each range
9.3 to 92.9999 V 1 to 20 kHz 0.5 0.25
20 to 100 kHz
[3]
5.0 0.5
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0% 0.5% Two digits on each range
93 to 929.999 mV 10 to 45 Hz 0.25 0.5
0.93 to 9.29999 V 45 Hz to 1 kHz 0.25 0.25 Six digits on each range
1 to 10 kHz 5.0 0.5
[1] To convert p-p to rms for trianglewave, multiply the p-p value by 0.2886751.
To convert p-p to rms for truncated sinewave, multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for truncated sine outputs is typical over this frequency band.
Squarewave
Ranges
p-p [1]
Frequency
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (% of output + % of range) [2]
Maximum
Voltage
Resolution
%Output %Range
Normal Channel (Single Output Mode)
2.9 to 65.999 mV 0.01 to 10 Hz 5.0% 0.5% Two digits on each range
66 to 659.999 mV 10 to 45 Hz 0.25 0.5
0.66 to 6.59999 V 45 Hz to 1 kHz 0.25 0.25 Six digits on each range
6.6 to 65.9999 V 1 to 20 kHz 0.5 0.25
20 to 100 kHz 5.0 0.5
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0% 0.5% Two digits on each range
66 to 659.999 mV 10 to 45 Hz 0.25 0.5
0.66 to 6.59999 V 45 Hz to 1 kHz 0.25 0.25 Six digits on each range
1 to 10 kHz 5.0 0.5
[1] To convert p-p to rms for squarewave, multiply the p-p value by .5000000.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.