5500A
Service Manual
1-26
1-21. AC Voltage (Sinewave) Extended Bandwidth Specifications
Ranges Frequency
1-Year Absolute
Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
Maximum Voltage
Resolution
%Output %Range
Normal Channel (Single Output Mode)
1.0 to 33 mV 0.01 to 10 Hz 5.0% 0.5% Two digits, e.g., 25 mV
34 to 330 mV Three digits
0.4 to 3.3 V Two digits
4 to 33 V Two digits
10 to 500 kHz (See AC Voltage (Sinewaves) Specifications)
0.3 to 3.3 V 500 kHz to 1
MHz
-8 dB at 1 MHz, typical
Two digits
1 to 2 MHz -32 dB at 2 MHz, typical
Auxiliary Output (Dual Output Mode)
10 to 330 mV 0.01 to 10 Hz 5.0% 0.5% Three digits
0.4 to 3.3 V Two digits
10 to 10 kHz (See AC Voltage (Sinewave) Specifications)

Table of Contents