1-27. AC Current (Non-Sinewave) Specifications
Trianglewave &
Truncated
Sinewave
Ranges [1]
Frequency
1-Year Absolute
Uncertainty, tcal ± 5 °C,
± (% of output + % of range)
[2]
Maximum
Current
Resolution
%Output %Range
2.9 to 92.999 mA 0.01 to 10 Hz 5.0% 0.5% Two digits, e.g., 75 mA
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.25 Six digits on each range
1 to 10 kHz 0.25 0.5
93 to 929.999 mA 0.01 to 10 Hz 5.0% 0.5% Two digits
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.5 Six digits on each range
1 to 10 kHz 5.0 1.0
0.93 to 2.19 A 10 to 45 Hz 5.0% 1.0% Two digits
45 Hz to 1 kHz 0.5 0.5 Six digits on each range
1 to 5 kHz 5.0 1.0
2.2 to 11 A 45 to 500 Hz 2.0% 0.5% Two digits on each range
500 Hz to 1 kHz 5.0 1.0 Six digits on each range
2.9 to 65.999 mA 0.01 to 10 Hz 5.0 % 0.5% Two digits, e.g., 50 mA
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.25 Six digits on each range
1 to 10 kHz 0.25 0.5
66 to 659.999 mA 0.01 to 10 Hz 5.0% 0.5% Two digits
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.5 Six digits on each range
1 to 10 kHz 5.0 1.0
0.66 to 2.19 A 10 to 45 Hz 5.0% 1.0% Two digits
45 Hz to 1 kHz 0.5 0.5 Six digits on each range
1 to 5 kHz 5.0 1.0
2.2 to 11 A 45 to 500 Hz 2.0% 0.5% Two digits on each range
500 Hz to 1 kHz 5.0 1.0 Six digits on each range
[1] All waveforms are p-p output ranges.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.