5500A
Service Manual
1-8
1-6. DC Current Specifications
Absolute Uncertainty, tcal ± 5 °C
Compliance Maximum
Ranges
± (% of output + µA)
Resolution Voltage Inductive
90 days 1 year
Load
0 to 3.29999 mA 0.010% 0.05 µA 0.013% 0.05 µA 0.01 µA 4.5 V 1 µH
0 to 32.9999 mA 0.008 0.25 0.01 0.25 0.1 4.5 V 200 µH
0 to 329.999 mA 0.008 3.3 0.01 3.3 1 4.5 to 3.0 V [1] 200 µH
0 to 2.19999 A 0.023 44 0.03 44 10 4.5 to 3.4 V [2] 200 µH
0 to 11 A 0.038 330 0.06 330 100 4.5 to 2.5 V [3] 200 µH
5725A Amplifier
0 to 11 A 0.03 330 0.04 330 100 4 V 400 µH
[1] The actual voltage compliance (V
c
) is a function of current output (I
o
), and is given by the formula:
V
c
= -5.05*I
o
+4.67. The highest compliance voltage is limited to 4.5 V.
[2] The actual voltage compliance (V
c
) is a function of current output (I
o
), and is given by the formula:
V
c
= -0.588*I
o
+4.69. The highest compliance voltage is limited to 4.5 V.
[3] The actual voltage compliance (V
c
) is a function of current output (I
o
), and is given by the formula:
V
c
= -0.204*I
o
+4.75. The highest compliance voltage is limited to 4.3 V.
Noise
Ranges
Bandwidth
0.1 to 10 Hz
Bandwidth
10 to 10 kHz
p-p rms
0 to 3.29999 mA 20 nA 200 nA
0 to 32.9999 mA 200 nA 2.0 µA
0 to 329.999 mA 2000 nA 20 µA
0 to 2.19999 A 20 µA1 mA
0 to 11 A 200 µA 10 mA
5725A Amplifier
0 to 11 A ± 25 ppm of output + 200 nA 2 mA