SC300 Option
Equipment Required for Calibration and Verification
6
6-75
6-99. Equipment Required for Calibration and Verification
Table 6-47 lists the equipment, recommended models, and minimum specifications
required for each calibration and verification procedure.
Table 6-47. SC300 Calibration and Verification Equipment
Instrument Model Minimum Use Specifications
Wave Generator, Edge Amplitude Calibration, AC Voltage Verification
Digital
Multimeter
HP 3458A
Voltage
1.8 mV to ± 105 V p-p Uncertainty: 0.06%
Edge 4.5 mV to 2.75 V p-p Uncertainty: 0.06%
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1% (used with Edge Amplitude
Calibration and AC Voltage Verification)
BNC Cable (supplied with SC300)
Edge Rise Time and Aberrations Verification
High-
Frequency
Digital Storage
Oscilloscope
Tektronix 11801 with
Tektronix SD-22/26
sampling head, or
Tektronix TDS 820 with
8 GHz bandwidth
Frequency 2 GHz
Resolution 4.5 mV to 2.75 V
Attenuator
Weinschel 9-10 (SMA)
or Weinschel 18W-10
or equivalent
10 dB, 3.5 mm (m/f)
Adapter
BNC(f) to 3.5 mm(m)
BNC Cable (supplied with SC300)
Leveled Sine Wave Amplitude Calibration and Verification
AC
Measurement
Fluke 5790A
Range 5 mV p-p to 5.5 V p-p
Standard
Frequency 50 kHz
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1%
BNC Cable (supplied with SC300)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Digital
Multimeter
HP 3458A
Adapter Pomona #1269
BNC(f) to Double Banana Plug
Termination Feedthrough 50 ± 1%
BNC Cable (supplied with SC300)