Table 6-41. SC300 Calibration and Verification Equipment (cont.)
Instrument Model Minimum Use Specifications
Leveled Sine Wave Frequency Verification
Frequency
Counter
PM 6680 with option (PM 9621, PM 9624, or
PM 9625) and (PM 9678)
50 kHz to 350 MHz, < 1.6 ppm
uncertainty
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC300)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
AC Measurement Fluke 5790A Range 5 mV p-p to 5.5 V p-p
Standard with -03 option Frequency 50 kHz to 10 MHz
Adapter
Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC300)
Leveled Sine Wave Harmonics Verification
Spectrum Analyzer HP 8590A
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC300)
Edge Frequency, AC Voltage Frequency Verification
Frequency Counter PM 6680 with option (PM
9678)
20 ms to 150 ns, 10 Hz to 10 MHz: < 1.6 ppm
uncertainty
BNC Cable (supplied with SC300)
Edge Duty Cycle
Frequency Counter PM 6680
BNC Cable (supplied with SC300)
Leveled Sine Wave Flatness (High Frequency) Calibration and Verification
Power Meter Hewlett-Packard
E4418A
Range -42 to +5.6 dBm
Frequency 10 - 300 MHz
Power Sensor Hewlett-Packard 8482A Range -20 to +19 dBm
Frequency 10 - 300 MHz
Power Sensor Hewlett-Packard 8481D Range -42 to -20 dBm
Frequency 10 - 300 MHz
30 dB
Reference
Hewlett-Packard
11708A
Range 30 dB
Attenuator (supplied with HP
8481D)
Frequency 50 MHz
Adapter Hewlett-Packard
PN 1250-1474
BNC(f) to Type N(f)
BNC Cable (supplied with SC300)