EasyManua.ls Logo

Fluke 5500A - AC Voltage (Sinewave) Specifications

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Service Manual
1-10
1-8. AC Voltage (Sinewave) Specifications
Ranges Frequency Absolute Uncertainty, tcal ± 5 °C
± (% of output + µV)
Res-
olution
Max
Burden
90 days 1 year [1]
1.0 to 32.999 mV 10 to 45 Hz 0.26% 20 µV 0.35% 20 µV1 µV 50
45 Hz to 10 kHz 0.11 20 0.15 20
10 to 20 kHz 0.15 20 0.2 20
20 to 50 kHz 0.19 20 0.25 20
50 to 100 kHz 0.26 33 0.35 33
100 to 500 kHz 0.75 60 1 60
33 to 329.999 mV 10 to 45 Hz 0.19 50 0.25 50 1 µV 50
45 Hz to 10 kHz 0.04 20 0.05 20
10 to 20 kHz 0.08 20 0.1 20
20 to 50 kHz 0.12 40 0.16 40
50 to 100 kHz 0.17 170 0.24 170
100 to 500 kHz 0.53 330 0.7 330
0.33 to 3.29999 V 10 to 45 Hz 0.11 250 0.15 250 10 µV 10 mA
45 Hz to 10 kHz 0.02 60 0.03 60
10 to 20 kHz 0.06 60 0.08 60
20 to 50 kHz 0.10 300 0.14 300
50 to 100 kHz 0.17 1700 0.24 1700
100 to 500 kHz 0.38 3300 0.5 3300
3.3 to 32.9999 V 10 to 45 Hz 0.11 2500 0.15 2500 100 µV 10 mA
45 Hz to 10 kHz 0.03 600 0.04 600
10 to 20 kHz 0.06 2600 0.08 2600
20 to 50 kHz 0.14 5000 0.19 5000
50 to 100 kHz 0.17 17000 0.24 17000
33 to 329.999 V 45 Hz to 1 kHz 0.04 6.6 mV 0.05 6.6 mV 1 mV 5 mA,
except
1 to 10 kHz 0.06 15 0.08 15 20 mA for
10 to 20 kHz 0.07 33 0.09 33 45 to 65 Hz
330 to 1020 V 45 Hz to 1 kHz 0.04 80 mV 0.05 80 mV 10 mV 2 mA,
except
1 to 5 kHz 0.15 100 0.20 100 6 mA for
5 to 10 kHz 0.15 500 0.20 500 45 to 65 Hz