Calibration and Verification
Performance Verification Tests
3
3-35
Table 3-30. Phase Accuracy Test (cont.)
Output
Voltage
(NORMAL)
Output
Voltage
(AUX)
Frequency Nominal
Phase
(degrees)
Measured
Value
(degrees)
Deviation % 1-Year Spec.
(degrees)
3 V 1 V 1 kHz 90 2
3 V 1 V 5 kHz 90 6
3 V 1 V 10 kHz 90 10
Output
Voltage
(NORMAL)
Output
Current
(AUX)
Frequency Nominal
Phase
(degrees)
Measured
Value
(degrees)
Deviation % 1-Year Spec.
(degrees)
33 V 300 mA 65 Hz 0 0.15
33 V 2 A 65 Hz 0 0.15
33 V 5 A 65 Hz 0 0.15
33 V 5 A 400 Hz 0 0.9
Table 3-31. Frequency Accuracy Test
Output
Voltage
(NORMAL)
Frequency Measured Value
(Hz)
Deviation ppm 1-Year Spec.
(ppm)
3 V 119.00 Hz 42
3 V 120.0 Hz 42
3 V 1000.0 Hz 27
3 V 100.00 kHz 25