5500A
Service Manual
1-20
1-14. AC Power (45 Hz to 65 Hz) Specification Summary, PF=1
5500A Calibrator Current Range
Voltage Range 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
5500A Calibrator
90 days 33 to 329.999 mV 0.30% 0.20% 0.25% 0.20%
330 mV to 1020 V 0.20 0.12 0.20 0.12
1 year 33 to 329.999 mV 0.40 0.25 0.35 0.25
330 mV to 1020 V 0.25 0.15 0.25 0.15
5725A Amplifier
90 days 100 to 1020 V 0.20 0.12 0.20 0.12
1 year 100 to 1020 V 0.25 0.15 0.25 0.15
5500A Calibrator Current Range
Voltage Range 0.33 to 0.8999 A 0.9 to 2.1999 A 2.2 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
5500A Calibrator
90 days 33 to 329.999 mV 0.25% 0.20% 0.25% 0.20%
330 mV to 1020 V 0.20 0.12 0.18 0.12
1 year 33 to 329.999 mV 0.35 0.25 0.35 0.25
330 mV to 1020 V 0.25 0.15 0.20 0.15
5725A Amplifier
90 days 100 to 1020 V 0.20 0.12 0.18 0.12
1 year 100 to 1020 V 0.25 0.15 0.20 0.15
5725A Amplifier Current Range
Voltage Range 1.5 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output)
5500A Calibrator
90 days 33 to 329.999 mV 0.25% 0.20%
330 mV to 1020 V 0.15 0.12
1 year 33 mV to 1020 V 0.35 0.25
330 mV to 1020 V 0.20 0.15
Note
1. To determine uncertainty with more precision, see “Calculating Power Uncertainty.”