Table 3-8. Resistance Calibration Steps
Step 5500A Output Comments
11 Ω Make a four-wire measurement
2 10 Ω “
3 11 Ω “
4 32 Ω “
5 35 Ω “
6 100 Ω “
7 110 Ω “
8 320 Ω “
9 0.35 kΩ “
10 1 kΩ “
11 1.1 kΩ “
12 3.2 kΩ “
13 3.5 kΩ “
14 10 kΩ “
15 11 kΩ “
16 32 kΩ “
17 35 kΩ “
18 100 kΩ “
19 110 kΩ Make a two-wire measurement
20 320 kΩ “
21 0.35 MΩ “
22 1 MΩ “
23 1.1 MΩ “
24 3.2 MΩ “
25 3.5 MΩ “
26 10 MΩ “
27 11 MΩ “
28 32 MΩ “
29 35 MΩ “
30 100 MΩ “
31 110 MΩ “