SC300 Option
SC300 Hardware Adjustments
6
6-111
Table 6-71. Wave Generator Verification at 50 Ω
Calibrator
Mainframe
Wave Type
Calibrator
Mainframe
output
(@ 10 kHz)
5790A
Reading
(V rms)
Conversion
Factor
5790A Reading x
Conversion Factor
(V p-p)
Tolerance
(V p-p)
square 5.0 mV 2.0000 250.00 µV
square 10.9 mV 2.0000 430.00 µV
square 45 mV 2.0000 1.450 mV
square 109 mV 2.0000 3.370 mV
square 0.45V 2.0000 13.570 mV
square 1.09V 2.0000 32.500 mV
square 2.20V 2.0000 66.100 mV
sine 5.0 mV 2.8284 250.00 µV
sine 10.9 mV 2.8284 430.00 µV
sine 45 mV 2.8284 1.450 mV
sine 109 mV 2.8284 3.370 mV
sine 0.45 V 2.8284 13.570 mV
sine 1.09 V 2.8284 32.500 mV
sine 2.20 V 2.8284 66.100 mV
triangle 5.0 mV 3.4641 250.00 µV
triangle 10.9 mV 3.4641 430.00 µV
triangle 45 mV 3.4641 1.450 mV
triangle 109 mV 3.4641 3.370 mV
triangle 0.45 V 3.4641 13.570 mV
triangle 1.09 V 3.4641 32.500 mV
triangle 2.20 V 3.4641 66.100 mV
6-136. SC300 Hardware Adjustments
Note
Before beginning SC300 hardware adjustments, it must determined which
revision of the option is installed in the instrument. To do this, remove the
top cover of the calibrator and look at the circuit board tab protruding
through the guard cover that is closest to the right front corner of the
calibrator. If this tab is marked A4, proceed to the“SC300 Hardware
Adjustments for the A4 Board” section of this manual.
Hardware adjustments must be made to the leveled sine and edge functions each time the
SC300 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.