5500A
Service Manual
6-110
4. Program the Calibrator Mainframe to output the wave type and voltage listed in
Table 6-71.
5. Allow the 5790A reading to stabilize, then record the 5790A rms reading for each
wave type and voltage in Table 6-71.
6. Multiply the rms reading by the conversion factor listed to convert it to the peak-to-
peak value.
7. Multiply the peak-to-peak value by (0.5 * (50 + Rload) / Rload), where Rload = the
actual feedthrough termination resistance, to correct for the resistance error.
Compare result to the tolerance column.
Table 6-70. Wave Generator Verification at 1 M
Calibrator
Mainframe
Wave Type
Calibrator
Mainframe
output
(@ 10 kHz)
5790A
Reading
(V rms)
Conversion
Factor
5790A Reading x
Conversion Factor
(V p-p)
Tolerance
(V p-p)
square 5.0 mV 2.0000 250.00 µV
square 20.0 mV 2.0000 700.00 µV
square 89 mV 2.0000 2.770 mV
square 219 mV 2.0000 6.670 mV
square 890 mV 2.0000 26.8 mV
square 6.5 V 2.0000 195.1 mV
square 55 V 2.0000 1.65 V
sine 5.0 mV 2.8284 250.00 µV
sine 20.0 mV 2.8284 700.00 µV
sine 89 mV 2.8284 2.770 mV
sine 219 mV 2.8284 6.670 mV
sine 890 mV 2.8284 26.8 mV
sine 6.5 V 2.8284 195.1 mV
sine 55 V 2.8284 1.65 V
triangle 5.0 mV 3.4641 250.00 µV
triangle 20.0 mV 3.4641 700.00 µV
triangle 89 mV 3.4641 2.770 mV
triangle 219 mV 3.4641 6.670 mV
triangle 890 mV 3.4641 26.8 mV
triangle 6.5 V 3.4641 195.1 mV
triangle 55 V 3.4641 1.65 V