Calibration and Verification
Performance Verification Tests
3
3-23
3-30. Resistance Accuracy
The Resistance Accuracy test verifies the accuracy of synthesized resistance at the
5500A Calibrator front panel NORMAL terminals. See Figure 3-3 for test equipment
connection instructions. For resistances of less than 110 k, use the four-wire COMP
option. For resistances of 110 k or higher, the COMP option is automatically turned
off. Table 3-16 shows the test points.
Table 3-16. Resistance Accuracy Test
Nominal Value Measured Value (Ohms) Deviation % 90-Day Spec. (m or %)
0 6 m
2 0.309%
10.9 0.064%
11.9 0.135%
19 0.088%
30 0.059%
33 0.052%
109 0.021%
119 0.020%
190 0.015%
300 0.012%
330 0.025%
1.09 k 0.012%
1.19 k 0.012%
1.9 k 0.010%
3 k 0.009%
3.3 k 0.025%
10.9 k 0.012%
11.9 k 0.012%
19 k 0.010%
30 k 0.009%
33 k 0.026%
109 k 0.013%
119 k 0.014%
190 k 0.012%