EasyManua.ls Logo

Fluke 5500A - High Frequency Flatness Verification at 5.5 V

Fluke 5500A
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC300 Option
Verification
6
6-103
3. Enter 10 MHz into the Calibrator Mainframe. Allow the power meter reading to
stabilize, then enter the power meter’s reading in Column B of Table 6-62.
4. Enter the next frequency listed in Table 6-62. Allow the power meter’s reading to
stabilize, then enter the reading into Column A of the table.
5. Enter 10 MHz into the Calibrator Mainframe. Allow the power meter reading to
stabilize, then enter the power meter’s reading in Column B of Table 6-62.
6. Repeat steps 4 and 5 for all of frequencies listed in Table 6-62. Continue until you
have completed Columns A and B.
7. When you have completed Columns A and B, press Y to remove the Calibrator
Mainframe’s output. Complete Table 6-62 by performing the calculations for each
column. Compare Column E to the specifications listed in the final column.
Table 6-62. High Frequency Flatness Verification at 5.5 V
Calibrator
Mainframe
Freq. (MHz)
A
B
10 MHz
CD E
Calibrator Mainframe
Flatness Spec. (%)
20 ± 1.50 +100 uV
50 ± 1.50 +100 uV
100 ± 1.50 +100 uV
125 ± 2.00 + 100 uV
160 ± 2.00 + 100 uV
200 ± 2.00 + 100 uV
220 ± 2.00 + 100 uV
235 ± 2.00 + 100 uV
250 ± 2.00 + 100 uV
300 ± 2.00 + 100 uV
Complete Columns A-E as follows:
A Enter the E4418A present frequency Reading (W).
B Enter the E4418A 10 MHz Reading (W).
C Apply power sensor correction factor for present frequency (W): CF * (Column A entry).
D Apply power sensor correction factor for 10 MHz (W): CF * (Column B entry).
E Compute and enter Error relative to 10 MHz (%): 100 * (sqrt(Column C entry) -
sqrt(Column D entry)) / sqrt(Column D entry).

Table of Contents

Other manuals for Fluke 5500A

Related product manuals