Calibration and Verification
Performance Verification Tests
3
3-25
3-32. AC Voltage Amplitude Accuracy (NORMAL)
The AC Voltage Amplitude Accuracy test verifies the accuracy of ac voltage at the
5500A Calibrator front panel NORMAL terminals. Table 3-18 shows the test points.
Table 3-18. AC Voltage Amplitude Accuracy Test (NORMAL)
Nominal Value Frequency Measured Value
(V)
(NORMAL)
Deviation % 90-Day Spec. (%)
30 mV 9.5 Hz 5.550
30 mV 10 Hz 0.327
30 mV 45 Hz 0.177
30 mV 1 kHz 0.177
30 mV 10 kHz 0.177
30 mV 20 kHz 0.217
30 mV 50 kHz 0.257
30 mV 100 kHz 0.370
30 mV 450 kHz 0.950
300 mV 9.5 Hz 5.550
300 mV 10 Hz 0.207
300 mV 45 Hz 0.047
300 mV 1 kHz 0.047
300 mV 10 kHz 0.047
300 mV 20 kHz 0.087
300 mV 50 kHz 0.133
300 mV 100 kHz 0.227
300 mV 500 kHz 0.640
3 V 9.5 Hz 5.550
3 V 10 Hz 0.118
3 V 45 Hz 0.022
3 V 1 kHz 0.022
3 V 10 kHz 0.022
3 V 20 kHz 0.062
3 V 50 kHz 0.110
3 V 100 kHz 0.227
3 V 450 kHz 0.490