Calibration and Verification
Calibration
3
3-5
3. You are then prompted to enter into the 5500A the value read on the measurement
instrument.
Note
Intermixed with these "output and measure" procedures are internal 5500A
calibration procedures that require no action by the operator.
3-6. DC Volts
Measure the 5500A output using a precision DMM, and enter into the 5500A each of the
measured values listed in Table 3-2 when prompted to do so.
Table 3-2. DC Volts Calibration Steps
Step 5500A Output (NORMAL)
1 +3.00 V
2 -3.00 V
3 +300 mV
4 +3.00 V
5 +30 V
6 +300 V
7 +1000 V
3-7. AC Volts
Measure the 5500A output using a precision ac voltmeter, and enter into the 5500A each
of the measured values listed in Table 3-3 when prompted to do so.
Table 3-3. AC Volts Calibration Steps
Step 5500A Output (NORMAL) Frequency
1 3.2999 V 100 Hz
2 0.330 V 100 Hz
3 3.00 V 500 kHz
4 3.00 V 9.99 Hz
5 30 mV 100 Hz
6 30 mV 500 kHz
7 300 mV 100 Hz
8 30 V 100 Hz
9 30 V 100 kHz
10 300 V 100 kHz
11 300 V 20 kHz
12 1000 V 100 Hz
13 1000 V 7 kHz