Table 6-31. Leveled Sine Wave Harmonics Verification
Calibrator Mainframe
Output Frequency
(@ 5.5 V p-p)
Harmonic HP 8590A Reading (dB) Tolerance
50 kHz 2 -33 dB
50 kHz 3, 4, 5 -46 dB
100 kHz 2 -33 dB
100 kHz 3, 4, 5 -38 dB
200 kHz 2 -33 dB
200 kHz 3, 4, 5 -38 dB
400 kHz 2 -33 dB
400 kHz 3, 4, 5 -38 dB
800 kHz 2 -33 dB
800 kHz 3, 4, 5 -38 dB
1 MHz 2 -33 dB
1 MHz 3, 4, 5 -38 dB
2 MHz 2 -33 dB
2 MHz 3, 4, 5 -38 dB
4 MHz 2 -33 dB
4 MHz 3, 4, 5 -38 dB
8 MHz 2 -33 dB
8 MHz 3, 4, 5 -38 dB
10 MHz 2 -33 dB
10 MHz 3, 4, 5 -38 dB
20 MHz 2 -33 dB
20 MHz 3, 4, 5 -38 dB
40 MHz 2 -33 dB
40 MHz 3, 4, 5 -38 dB
80 MHz 2 -33 dB
80 MHz 3, 4, 5 -38 dB
100 MHz 2 -33 dB
100 MHz 3, 4, 5 -38 dB
200 MHz 2 -33 dB
200 MHz 3, 4, 5 -38 dB
400 MHz 2 -33 dB
400 MHz 3, 4, 5 -38 dB
600 MHz 2 -33 dB
600 MHz 3, 4, 5 -38 dB