6-50. AC Voltage Frequency Verification
This procedure uses the following equipment:
PM 6680 Frequency Counter with an ovenized timebase (Option PM 9690 or PM
9691)
BNC cable supplied with the SC600
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
At 50 MHZ
SC600 Cable
PM 6680A
5500A-SC600
om057f.eps
Figure 6-6. AC Voltage Frequency Verification Setup
Set the Calibrator Mainframe to SCOPE mode, with the Volt menu on the display. Press
O on the Calibrator Mainframe to activate the output. Then follow these steps to
verify AC Voltage frequency.
1. Set the PM 6680’s FUNCTION to measure frequency on channel A with auto
trigger, measurement time set to 1 second or longer, 1M impedance, and filter off.
2. Using the BNC cable, connect the SCOPE connector on the Calibrator Mainframe to
PM 6680 channel A.
3. Program the Calibrator Mainframe to output 2.1 V at each frequency listed in Table
6-23.
4. Allow the PM 6680 reading to stabilize, then record the PM 6680 reading for each
frequency listed in Table 6-23. Compare to the tolerance column of Table 6-23.
Table 6-23. AC Voltage Frequency Verification
Calibrator Mainframe
Frequency
(output @ 2.1 V p-p)
PM 6680 Reading
(Frequency) Tolerance
10 Hz 0.000025 Hz
100 Hz 0.00025 Hz
1 kHz 0.0025 Hz
10 kHz 0.025 Hz