5500A
Service Manual
1-16
1-11. Temperatore Calibration (Thermocouple) Specifications
TC
Type
Range (°C) Absolute
Uncertainty
Source/Measure,
tcal ± 5 °C, ± (°C)
[1]
TC
Type
Range (°C) Absolute
Uncertainty
Source/Measure,
tcal ± 5 °C, ± (°C)
[1]
90 days 1 year 90 days 1 year
B 600 °C to 800 °C 0.42 °C 0.44 °C L -200 to -100 0.37 0.37
800 to 1000 0.34 0.34 -100 to 800 0.26 0.26
1000 to 1550 0.30 0.30 800 to 900 0.17 0.17
1550 to 1820 0.26 0.33 N -200 to -100 0.30 0.40
C 0 to 150 0.23 0.30 -100 to -25 0.17 0.22
150 to 650 0.19 0.26 -25 to 120 0.15 0.19
650 to 1000 0.23 0.31 120 to 410 0.14 0.18
1000 to 1800 0.38 0.50 410 to 1300 0.21 0.27
1800 to 2316 0.63 0.84 R 0 to 250 0.48 0.57
E -250 to -100 0.38 0.50 250 to 400 0.28 0.35
-100 to -25 0.12 0.16 400 to 1000 0.26 0.33
-25 to 350 0.10 0.14 1000 to 1767 0.30 0.40
350 to 650 0.12 0.16 S 0 to 250 0.47 0.47
650 to 1000 0.16 0.21 250 to 1000 0.30 0.36
J -210 to -100 0.20 0.27 1000 to 1400 0.28 0.37
-100 to -30 0.12 0.16 1400 to 1767 0.34 0.46
-30 to 150 0.10 0.14 T -250 to -150 0.48 0.63
150 to 760 0.13 0.17 -150 to 0 0.18 0.24
760 to 1200 0.18 0.23 0 to 120 0.12 0.16
K -200 to -100 0.25 0.33 120 to 400 0.10 0.14
-100 to -25 0.14 0.18 U -200 to 0 0.56 0.56
-25 to 120 0.12 0.16 0 to 600 0.27 0.27
120 to 1000 0.19 0.26
1000 to 1372 0.30 0.40
The 10 µV/°C linear output mode has the same uncertainty as the 300 mV dc range.
Applies to both simulated thermocouple output and thermocouple measurement.
[1] Temperature standard ITS-90 or IPTS-68 is selectable.
[2] Resolution is 0.01 °C.
[3] Does not include thermocouple error.