Calibration and Verification
Calibration
3
3-11
Table 3-8. Resistance Calibration Steps (cont.)
Step 5500A Output Comments
32 320 M [1] Make a two-wire measurement
[1] Perform this test using the HP 3458A in the 10 M range and the Fluke 742A-10M in
parallel with the 5500A output. Using exactly 10 M, the nominal value displayed on the HP 3458A is
9.66667 M. Figure 3-4 shows the connections and the equation you use to calculate actual
resistance. Enter the calculated actual resistance, R
UUT
, into the HP 3458A. In the equation, R
3458
is the
reading of the HP 3458A, R
742
is the printed value of the 742A-10M, and R
UUT
is the actual 5500A
output.
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
5500A
HP3458 4W Ohms
Function
742A-10M
R
UUT
=
R
3458
(
(
R
742
(
(
R
742
R
3458
_
om011f.eps
Figure 3-4. High End Resistance Connections with Equation