5500A Phase Specifications (cont.)
Phase
(Φ)
Phase
(Φ)
PF Power Uncertainty Adder due to Phase Error
Watts VARs 10 to 65 Hz 65 to 500 Hz 500 to 1 kHz 1 to 5 kHz 5k to
10 kHz
0 degrees 90 degrees 1.000 0.00% 0.01% 0.06% 0.55% 1.52%
5 85 0.996 0.02 0.15
10 80 0.985 0.05 0.29
15 75 0.966 0.07 0.43
20 70 0.940 0.10 0.58
25 65 0.906 0.12 0.74
30 60 0.866 0.15 0.92
35 55 0.819 0.18 1.11
40 50 0.766 0.22 1.33
45 45 0.707 0.26 1.58
50 40 0.643 0.31 1.88
55 35 0.574 0.37 2.26
60 30 0.500 0.45 2.73
65 25 0.423 0.56 3.38
70 20 0.342 0.72 4.33
75 15 0.259 0.98 5.87
80 10 0.174 1.49 8.92
85 5 0.087 2.99 17.97
90 0 0.000
To calculate exact ac watts power adders due to phase uncertainty for values not shown, use the
following formula
:
()
()
Adder
Cos
Cos
%(
()
=−
+
100 1
Φ
Φ
)
. For example: for
a PF of .9205
(Φ = 23) and a phase uncertainty of
∆Φ
= 0.15, the ac watts power adder is:
()
()
Adder
Cos
Cos
%(
(23 )
=−
+
=100 1
15
23
011%
.
).