5500A
Service Manual
1-18
Temperature Calibration (RTD) Specifications (cont.)
Pt 385, 500 Ω -200 -80 0.03 0.04
-80 0 0.04 0.05
0 100 0.05 0.05
100 260 0.06 0.06
260 300 0.07 0.08
300 400 0.07 0.08
400 600 0.08 0.09
600 630 0.09 0.11
Pt 385, 1000 Ω -200 -80 0.03 0.03
-80 0 0.03 0.03
0 100 0.03 0.04
100 260 0.04 0.05
260 300 0.05 0.06
300 400 0.05 0.07
400 600 0.06 0.07
600 630 0.22 0.23
PtNi 385, 120 Ω -80 0 0.06 0.08
(Ni120) 0 100 0.07 0.08
100 260 0.13 0.14
Cu 427, 10 Ω [3] -100 260 0.3 0.3
[1] Resolution is 0.003 °C.
[2] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire
compensation.
[3] Based on MINCO Application Aid No. 18.