5500A
Service Manual
1-12
AC Voltage (Sinewave) Specifications (cont.)
Ranges Frequency
Maximum Distortion and Noise
10 Hz to 5 MHz Bandwidth
± (% output + µV)
1.0 to 32.999 mV 10 to 45 Hz 0.15% + 90 µV
45 Hz to 10 kHz 0.035 + 90
10 to 20 kHz 0.06 + 90
20 to 50 kHz 0.15 + 90
50 to 100 kHz 0.25 + 90
100 to 500 kHz 0.3 + 90
33 to 329.999 mV 10 to 45 Hz 0.15% + 90 µV
45 Hz to 10 kHz 0.035 + 90
10 to 20 kHz 0.06 + 90
20 to 50 kHz 0.15 + 90
50 to 100 kHz 0.20 + 90
100 to 500 kHz 0.20 + 90
0.33 to 3.29999 V 10 to 45 Hz 0.15% + 200 µV
45 Hz to 10 kHz 0.035 + 200
10 to 20 kHz 0.06 + 200
20 to 50 kHz 0.15 + 200
50 to 100 kHz 0.20 + 200
100 to 500 kHz 0.20 + 200
3.3 to 32.9999 V 10 to 45 Hz 0.15% + 2 mV
45 Hz to 10 kHz 0.035 + 2
10 to 20 kHz 0.08 + 2
20 to 50 kHz 0.2 + 2
50 to 100 kHz 0.5 + 2
33 to 329.999 V 45 Hz to 1 kHz 0.15% + 10 mV
1 to 10 kHz 0.05 + 10
10 to 20 kHz 0.6 + 10
330 to 1000 V 45 Hz to 1 kHz 0.15% + 30 mV
1 to 10 kHz 0.07 + 30
5725A Amplifier
100 to 1000 V 45 Hz to 1 kHz 0.07%
1 to 20 kHz 0.15%
20 to 30 kHz 0.3%
100 to 750 V 30 to 100 kHz 0.4%
Auxiliary Output [dual output mode only] 10 Hz to 100 kHz Bandwidth
10 to 329.999 mV 10 to 20 Hz 0.2% + 200 µV
20 to 45 Hz 0.06 + 200
45 Hz to 1 kHz 0.08 + 200
1 to 5 kHz 0.3 + 200
5 to 10 kHz 0.6 + 200
0.33 to 3.29999 V 10 to 20 Hz 0.2% + 200 µV
20 to 45 Hz 0.06 + 200
45 Hz to 1 kHz 0.08 + 200
1 to 5 kHz 0.3 + 200
5 to 10 kHz 0.6 + 200