SC600 Option
Verification
6
6-45
Connect the HP E4418A Power Meter to either the 8482A or the 8481D Power Sensor as
shown in Figure 6-11. For more information on connecting the two instruments, see the
power meter and power sensor operators manuals.
Connect the power meter/power sensor combination to the SCOPE connector on the
Calibrator Mainframe, as shown in Figure 6-12.
The Hewlett-Packard E4418A Power Meter must be configured by setting the
parameters listed below. Zero and self-calibrate the power meter with the power sensor
being used. Refer to the Hewlett-Packard E4418A operators manual for details.
PRESET
RESOLN 3
AUTO FILTER
WATTS
SENSOR TABLE 0 (default)
om035f.eps
Figure 6-11. Connecting the HP E4418A Power Meter to the HP 8482A or 8481D Power Sensor
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
om036f.eps
Figure 6-12. Connecting the Calibrator Mainframe to the HP Power Meter and Power Sensor