Table 6-51. AC Voltage Verification at 1 M
Nominal Value (p-p) Frequency Measured Value (p-p) Deviation (mV) 1-Year Spec. (mV)
5.0 mV 10 Hz 0.11
5.0 mV 100 Hz 0.11
5.0 mV 1 kHz 0.11
5.0 mV 5 kHz 0.11
5.0 mV 10 kHz 0.11
10.0 mV 10 kHz 0.12
20.0 mV 100 Hz 0.15
20.0 mV 1 kHz 0.15
20.0 mV 10 kHz 0.15
50.0 mV 10 kHz 0.23
89.0 mV 10 Hz 0.32
89.0 mV 10 kHz 0.32
100.0 mV 10 kHz 0.35
200.0 mV 100 Hz 0.60
200.0 mV 1 kHz 0.60
200.0 mV 10 kHz 0.60
500.0 mV 10 kHz 1.35
890.0 mV 10 Hz 2.32
890.0 mV 10 kHz 2.32
1.0 V 100 Hz 2.60
1.0 V 1 kHz 2.60
1.0 V 10 kHz 2.60
2.0 V 10 kHz 5.10
5.0 V 10 Hz 12.60
5.0 V 10 kHz 12.60
10.0 V 10 kHz 25.10
20.0 V 10 kHz 50.10
50.0 V 10 Hz 125.10
50.0 V 100 Hz 125.10
50.0 V 1 kHz 125.10
50.0 V 10 kHz 125.10
105.0 V 100 Hz 262.60
105.0 V 1 kHz 262.60

Table of Contents