EasyManuals Logo

Fluke 5700A Series II User Manual

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
5700A/5720A Series II
Operators Manual
4-14
Caution
Use only cables with correct voltage ratings.
Table 4-1. Auxiliary Amplifier Data
Output Function Cable Recommendations
DC voltage
AC voltage 10 kHz
AC current 2A, 10 kHz
DC current 2A
Resistance
Low Thermal EMF Test Leads [Note 1]
5440A-7002 (banana plugs) or
5440A-7003 (spade lugs) 2 ft. (61 cm.) cable
(If external sensing is required, use a twisted
shielded pair.)
AC voltage >10 kHz SENSE/GUARD: Triaxial cable or Twinax (e.g.,
Alpha 2829/2), OUTPUT: Coaxial
Or: SENSE: Coaxial, OUTPUT: Coaxial
GUARD Lead: Separate wire
AC current with guard Triaxial cable
Wideband AC 6-foot (2 m) 50 coaxial cable with type “N” male
connector supplied with the option. A 50
feedthrough terminator is also supplied for
connecting to meters with an impedance >50.
Voltage-boosted output, 5725A Low Thermal EMF Test Leads [Note 1]
5440A-7002 (banana plugs) or
5440A-7003 (spade lugs)
(Output is at the calibrator’s front panel.)
Current-boosted output, 5725A 16-gauge or heavier twisted pair insulated wire, as
short as possible to minimize resistance and
inductance. (Output is at the amplifier terminals.)
1. Spade lugs provide a slightly better thermal EMF performance. However, some UUTs have repressed
banana connectors that cannot accommodate spade lugs.
4-19. When to Use External Sensing
External sensing is normally required only when you are calibrating a device that draws
enough current to produce a significant voltage drop in the cables. An example of such a
case is using the calibrator as an external dc voltage reference for an AC/DC transfer
standard. In this example, the calibrator is sourcing 1V dc into a Fluke 540B AC/DC
Transfer Standard. The 180 input impedance results in a current flow of approximately
5 mA. The calibrator’s 90-day uncertainty at 1V is specified to be ±(6 ppm + 1.2 uV) or ±
7.2 µV. Cumulative lead and contact resistances of as little as 2 m would cause a
voltage drop greater than the total uncertainty of the calibrator. External sensing
eliminates this error.
The normal power-up state of the calibrator is external sensing off, with an internal
connection between the SENSE and OUTPUT automatically made. This is the state
achieved by pressing X so that the indicator is off.

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals