5700A/5720A Series II
Operators Manual
1-36
1-33. AC Current Specifications
5720A Series II AC Current Specifications: 99 % Confidence Level
Absolute Uncertainty
±5 °C from calibration temperature
[1]
Relative Uncertainty
±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
Range Resolution
Frequency
(Hz)
± (ppm output + nA)
220 µA
1 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
260 + 20
170 + 12
120 + 10
300 + 15
1000 + 80
280 + 20
180 + 12
130 + 10
320 + 15
1100 + 80
290 + 20
190 + 12
135 + 10
340 + 15
1200 + 80
300 + 20
200 + 12
140 + 10
350 + 15
1300 + 80
260 + 20
130 + 12
100 + 10
250 + 15
900 + 80
280 + 20
150 + 12
110 + 10
280 + 15
1000 + 80
2.2 mA 10 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
260 + 50
170 + 40
120 + 40
210 + 130
1000 + 800
280 + 50
180 + 40
130 + 40
220 + 130
1100 + 800
290 + 50
190 + 40
135 + 40
230 + 130
1200 + 800
300 + 50
200 + 40
140 + 40
240 + 130
1300 + 800
260 + 50
130 + 40
100 + 40
250 + 130
900 + 800
280 + 50
150 + 40
110 + 40
280 + 130
1000 + 800
22 mA 100 nA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
260 + 500
170 + 400
120 + 400
210 + 700
1000 + 6000
280 + 500
180 + 400
130 + 400
220 + 700
1100 + 6000
290 + 500
190 + 400
135 + 400
230 + 700
1200 + 6000
300 + 500
200 + 400
140 + 400
240 + 700
1300 + 6000
260 + 500
130 + 400
100 + 400
250 + 700
900 + 6000
280 + 500
150 + 400
110 + 400
280 + 700
1000 + 6000
± (ppm output + µA)
220 mA
1 µA
10 – 20
20 – 40
40 – 1 k
1k – 5 k
5k – 10 k
260 + 5
170 + 4
120 + 3
210 + 4
1000 + 12
280 + 5
180 + 4
130 + 3
220 + 4
1100 + 12
290 + 5
190 + 4
135 + 3
230 + 4
1200 + 12
300 + 5
200 + 4
140 + 3
240 + 4
1300 + 12
260 + 5
130 + 4
100 + 3
250 + 4
900 + 12
280 + 5
150 + 4
110 + 3
280 + 4
1000 + 12
2.2 A
10 µA
20 – 1 k
1 k – 5 k
5 k – 10 k
290 + 40
440 + 100
6000 + 200
300 + 40
460 + 100
7000 + 200
310 + 40
480 + 100
7500 + 200
320 + 40
500 + 100
8000 + 200
300 + 40
500 + 100
6000 + 200
350 + 40
520 + 100
7000 + 200
5725A Amplifier:
11 A
100 µA
40 – 1 k
1 k – 5 k
5 k – 10 k
370 + 170
800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
440 + 170
900 + 380
3500 + 750
460 + 170
950 + 380
3600 + 750
300 + 170
700 + 380
2800 + 750
330 + 170
800 + 380
3200 + 750
Note:
Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 µA and 2.2 mA ranges are increased by a factor
of 1.3 plus 2 µA when supplied through 5725A terminals. For the 5720A 220 µA range, 1 kHz through 5 kHz and 5 kHz through 10 kHz,
when the output is coming from the AUX current terminal, use the 5700A Absolute Uncertainty Specifications. Specifications are otherwise
identical for all output locations.
1. For fields strengths >0.4 V/m but 3 V/m, add 1 % of range.

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V