5700A/5720A Series II
Operators Manual
5-54
5-39. IEEE-488 Interface Configuration
The calibrator’s IEEE-488 interface supports the IEEE-488 interface function subsets
listed in Table 5-7.
Table 5-7. Supported IEEE-488 Interface Function Subsets
Interface Function Description
SH1 Complete source handshake capability
AH1 Complete acceptor handshake capability
T6 Basic talker; serial poll; no talk-only mode; unaddress if MLA
TEO No extended talker capabilities
L4 Basic listener operation; no listen-only mode; unaddress if MTA
LEO No extended listener capabilities
SR1 Full service request capability with ability to bit-mask SRQ
RL1 Full remoter/local capability including local lockout
PPO No parallel poll capability
DC1 Device clear capability
DT0 No device trigger capability
C0 No bus control capability

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals