Remote Operation
Remote Program Examples 5
5-65
5-52. Wideband AC Module (Option 5700A-03) Gain Calibration
! THIS IS A PROGRAM FOR REMOTE WIDEBAND (OPTION -Ø3) GAIN CALIBRATION
! WHICH IS PART OF ROUTINE CALIBRATION IF A WIDEBAND MODULE IS INSTALLED
3Ø !
4Ø REMOTE
PRINT @4, “*CLS ; *SRE 8” ! ENABLE FAULT TO TRIGGER SRQ
ON SRQ GOTO 8Ø ! SRQ HANDLER
GOTO 15Ø ! START MAIN PROGRAM
PRINT @4, “FAULT?” \ INPUT @4, A% ! SRQ HANDLER, GET FAULT CODE
IF A% = Ø GOTO 13Ø ! IF NO FAULT EXIT
1ØØ PRINT @4, “EXPLAIN? “;A% \ INPUT @4, A$ ! GET EXPLANATION
11Ø PRINT “FAULT”;A%;”OCCURRED: “;A$ ! DISPLAY EXPLANATION
12Ø GOTO 8Ø ! GET NEXT FAULT
13Ø A% = SPL(4%) ! SERIAL POLL TO CLEAR SP BYTE
14Ø RESUME 42Ø
15Ø PRINT “Ambient temperature in Celsius”; ! MAIN PROGRAM BEGINS
16Ø INPUT T ! GET TEMPERATURE FROM OPERATOR
17Ø PRINT @4, “CAL_TEMP”;T ! SET CAL TEMPERATURE
18Ø PRINT “Connect the wideband cable to the SENSE binding posts through the”
19Ø PRINT “5Ø ohm terminator with shield to LO as shown in the manual.”
2ØØ GOSUB 37Ø ! PAUSE FOR KEY
21Ø PRINT “Calibrating wideband positive gain, please wait ...”
22Ø PRINT @4, “CAL_WBGAIN PGAIN;*OPC” \ GOSUB 39Ø ! CALIBRATES POSITIVE GAIN
23Ø PRINT “Reverse the polarity at the SENSE binding posts.”
24Ø GOSUB 37Ø ! PAUSE FOR KEY
25Ø PRINT “Calibrating wideband negative gain, please wait ...”
26Ø PRINT @4, “CAL_WBGAIN NGAIN;*OPC” \ GOSUB 39Ø! CALIBRATES NEGATIVE GAIN
27Ø PRINT “Wideband gains calibration finished. Print a report? [y, n]”
28Ø INPUT A$ \ IF LEFT(A$,1)=“N” OR LEFT(A$,1)=“n” GOTO 31Ø
29Ø PRINT “Set up the peripheral device.” \ GOSUB 37Ø
3ØØ PRINT “Printing . . .” \ PRINT @4, “CAL_PR CHECK;*OPC” \ GOSUB 39Ø
31Ø PRINT “Store or discard the constants? [s, d]” \ INPUT A$
32Ø IF LEFT(A$,1)=“D” OR LEFT(A$,1)=“d” GOTO 42Ø
33Ø PRINT “Set the CALIBRATION switch to ENABLE” \ GOSUB 37Ø
34Ø PRINT “Saving the constants in memory . . .”
35Ø PRINT @4,”CAL_WBGAIN STORE;*OPC” \ GOSUB 39Ø ! SAVE THE CONSTANTS
36Ø PRINT “Set the CALIBRATION switch to NORMAL” \ GOSUB 37Ø \ GOTO 42Ø
37Ø PRINT “Press RETURN to continue” ! PAUSE FOR OPERATOR SUBROUTINE
38Ø INPUT A$ \ RETURN ! “
39Ø PRINT @4, “*ESR?”\ INPUT @4,A% ! WAIT FOR OPC SUBROUTINE
4ØØ IF A% AND 1% THEN RETURN ! “, RETURN IF OPC BIT IS 1
41Ø GOTO 39Ø ! “, STAY IN LOOP IF OPC IS Ø
42Ø PRINT “All finished!” \ END

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals