Using the RS-232C Serial Interface
Printing Calibration Reports 6
6-7
6-6. Calibration Check Shift Results
The Calibration Check Shift Results printout present changes in the magnitude of self
correction generated by calibration check. Calibration check uses internal standards
rather than external standards as the basis for comparison, and does not make permanent
changes to calibration constants. Chapter 1 describes how to use calibration check to
develop a performance history. The printout contains the following information:
• Heading that shows the date, time, and user report string
• Modules present (installed or attached)
• Most recent calibration dates and temperatures
• Values of externally-calibrated internal references
• DC voltage output shifts for each range
• AC voltage output shifts for each range
• AC and dc current output shifts for each range
• Resistance shifts for each value
• 5725A Boost Amplifier output shifts (if attached)