Using the RS-232C Serial Interface
Printing Calibration Reports 6
6-7
6-6. Calibration Check Shift Results
The Calibration Check Shift Results printout present changes in the magnitude of self
correction generated by calibration check. Calibration check uses internal standards
rather than external standards as the basis for comparison, and does not make permanent
changes to calibration constants. Chapter 1 describes how to use calibration check to
develop a performance history. The printout contains the following information:
Heading that shows the date, time, and user report string
Modules present (installed or attached)
Most recent calibration dates and temperatures
Values of externally-calibrated internal references
DC voltage output shifts for each range
AC voltage output shifts for each range
AC and dc current output shifts for each range
Resistance shifts for each value
5725A Boost Amplifier output shifts (if attached)