EasyManua.ls Logo

Fluke 5700A Series II - Linearity Checking Using Offset and Scale

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Front Panel Operation
Linearity Checking Using Offset and Scale 4
4-47
4-50. Linearity Checking Using Offset and Scale
Using the calibrator’s offset and scaling features, you can remove a UUT's offset and
scale errors to isolate and display linearity error. The following procedure is an example
that uses
o and S to determine both scale and linearity error of a 4 1/2-digit DMM.
In this example, the DMM is set to the 20V dc range, and the calibrator is connected to
the DMM. Figure 4-10 illustrates all three types of errors that are detected by the
calibrator. The numbers in the graph correspond to the conditions encountered in the
example.
1.
Set the calibrator to 0 mV in standby. The displays change to:
Output Display
Control Display
UUT (DMM) Display
2.
Use the output adjustment controls (the rotary knob and arrow keys) to adjust the
calibrator’s output for a reading of 0V on the DMM. The displays change to:
Output Display
Control Display
UUT (DMM) Display
CALIBRATOR
VOLTAGE
0
-1.3 mV
METER READING
OFFSET
ERROR =
+1.3 mV
10.000
10.007
19.900
19.9017
19.8987
10.000208
9.993208
LINEAR
RESPONSE
ACTUAL
METER
RESPONSE
LINEARITY
ERROR
SCALE
ERROR
19.9
IDEAL
RESPONSE
Figure 4-10. Meter Response vs. Stimulus

Table of Contents

Other manuals for Fluke 5700A Series II

Related product manuals