EasyManuals Logo

Fluke 5700A Series II User Manual

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
Remote Operation
Remote Program Examples 5
5-63
5-50. Calibration
! THIS IS A PROGRAM FOR REMOTE MAIN OUTPUT CALIBRATION
2Ø !
25 REMOTE
PRINT @4, “*CLS; *SRE 8” ! ENABLE EAV TO TRIGGER SRQ
ON SRQ GOTO 6Ø ! SRQ HANDLER
GOTO 14Ø ! START MAIN PROGRAM
PRINT @4, “FAULT?” \ INPUT @4, A% ! SRQ HANDLER, GET FAULT CODE
IF A% = Ø GOTO 12Ø ! IF NO FAULT EXIT
PRINT @4, “EXPLAIN? “;A% \ INPUT @4, A$ ! GET EXPLANATION
9Ø PRINT
1ØØ PRINT “FAULT”;A%;”OCCURRED: “;A$ ! DISPLAY EXPLANATION
11Ø GOTO 6Ø! GET NEXT FAULT
12Ø A% = SPL(4%) ! SERIAL POLL TO CLEAR SP BYTE
13Ø RESUME 53Ø ! END OF SRQ HANDLER
14Ø PRINT “Ambient temperature in Celsius”; ! MAIN PROGRAM BEGINS
15Ø INPUT T ! GET TEMPERATURE FROM OPERATOR
16Ø PRINT @4, “CAL_TEMP”;T ! SET CAL TEMPERATURE
17Ø PRINT “Connect the 1ØV standard as shown in the manual” \ GOSUB 47Ø
18Ø PRINT “Enter the true value of 1ØV standard in volts”; \ INPUT V
19Ø PRINT “Doing 1ØV reference calibration. Please wait ...”
2ØØ PRINT @4, “CAL_REF “;V;”V;*OPC” \ GOSUB 49Ø
21Ø PRINT “Reverse the polarity of the leads” \ GOSUB 47Ø
22Ø PRINT “Doing -1ØV reference calibration. Please wait ...”
23Ø V = -V ! SET THE VARIABLE FOR -1ØV CAL
24Ø PRINT @4, “CAL_REF “;V;”V;*OPC” \ GOSUB 49Ø
25Ø PRINT “Connect the 1Ø kilohm standard to the output terminals”\GOSUB 47Ø
26Ø PRINT “Enter the true value of the 1Ø kilohm standard in kilohms”;
27Ø INPUT K1
28Ø PRINT “Doing 1Ø kilohm reference calibration. Please wait ...”
29Ø PRINT @4, “CAL_REF “;K1;”KOHM;*OPC” \ GOSUB 49Ø
3ØØ PRINT “Connect the 1 ohm standard to the output terminals”\GOSUB 47Ø
31Ø PRINT “Enter the true value of the 1 ohm standard in ohms”; \ INPUT K2
32Ø PRINT “Doing 1 ohm reference calibration. Please wait ...”
33Ø PRINT @4, “CAL_REF “;K2;”OHM;*OPC” \ GOSUB 49Ø
34Ø PRINT “Doing internal portion of calibration. Please wait ...”
35Ø PRINT @4, “CAL_ADJ;*OPC” \ GOSUB 49Ø
36Ø PRINT “Calibration finished. Print a calibration report? [y, n]”
37Ø INPUT A$ \ IF LEFT(A$,1)=“N” OR LEFT(A$,1)=“n” GOTO 4ØØ
38Ø PRINT “Set up the peripheral device.” \ GOSUB 47Ø
39Ø PRINT “Printing . . .” \ PRINT @4, “CAL_PR CHECK;*OPC” \ GOSUB 49Ø
4ØØ PRINT “Store or discard the constants? [s, d]” \ INPUT A$
41Ø IF LEFT(A$,1)=“D” OR LEFT(A$,1)=“d” GOTO 53Ø
42Ø PRINT “Set the CALIBRATION switch to ENABLE” \ GOSUB 47Ø
43Ø PRINT “Saving the constants in memory . . .”
44Ø PRINT @4, “CAL_STORE;*OPC” \ GOSUB 49Ø ! SAVE THE CONSTANTS IN MEMORY
45Ø PRINT “Set the CALIBRATION switch to NORMAL” \ GOSUB 47Ø \ GOTO 53Ø
47Ø PRINT “Press RETURN to continue” ! PAUSE FOR OPERATOR SUBROUTINE
48Ø INPUT A$ \ RETURN ! “
49Ø PRINT @4, “*ESR?” \ INPUT @4, A% ! WAIT FOR OPC SUBROUTINE
51Ø IF A% AND 1% THEN RETURN ! “, RETURN IF OPC BIT IS 1
52Ø GOTO 49Ø ! “, STAY IN LOOP IF OPC IS Ø
53Ø PRINT “All finished!” \END

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals