Introduction and Specifications
Support Equipment and Services 1
1-7
1-10. 732B-200 Direct Volt Maintenance Program (U.S.A. Only)
The Fluke 732B-200 Direct Volt Maintenance Program provides your laboratory with
NIST-traceable 10 V calibration uncertainty as low as 0.6 parts per million.
The program maintains the 732B that you keep in your laboratory. To accomplish this,
the following occurs:
1. Fluke sends you a calibrated Fluke-owned 732B standard, together with all-necessary
connecting cables and instructions for comparison with your 10 V reference standard.
2. You take a series of readings over a five-day period, and return the results to the
Fluke Standards Laboratory.
3. The Fluke Standards Laboratory assigns a value to your 10 V standard relative to the
NIST legal volt and sends you a report of calibration.
1-11. 742A Series Resistance Standards
The calibrator uses 1 and 10 k resistor standards such as the 742A Series in its semi-
automated calibration procedure to establish external traceability of resistance and
current. Chapter 7 describes this procedure.
The 742A Resistance Standards, which are constructed of arrays of Fluke wirewound
precision resistors, are ideally suited as support standards for the calibrator. Stability of
the resistance transfer standards and their temperature coefficients make them ideal for
easy transport to and operation in the calibrator's working environment.
1-12. Wideband AC Module (Option 5700A-03) Calibration Support
The Wideband AC Module (Option 5700A-03) requires two kinds of calibration: gain
and flatness. Gain constants are checked and recalibrated as a part of the normal
calibrator semi-automated calibration process.
Since frequency flatness is determined by such stable parameters as circuit geometry and
dielectric constants, flatness of the Wideband AC module has excellent long-term
stability. This stability gives the Wideband AC Module a two-year calibration cycle for
flatness calibration. Flatness calibration is required only infrequently, and can be done
when the calibrator is returned to a standards laboratory for periodic verification. The
5700A/5720A Series II Service Manual contains the wideband flatness calibration
procedure. Chapter 7 of this manual contains the wideband gain calibration procedure.
1-13. Service Centers
A worldwide network of Fluke service centers supports Fluke instruments and assists
customers in many ways. Most service centers have standards and calibration laboratories
certified by local national standards organizations. The following is a partial list of the
services provided by most service centers:
Repair and certified traceable calibration of all Fluke products.
Certified traceable calibration of many non-Fluke standards and calibrators.
Worldwide exchange of calibrator internal modules. Delivery inside the U.S.A. is
typically within 48 hours.
Service agreements with the flexibility to suit your needs. These can be a simple
warranty extension or an agreement that includes on-site support. Calibration service
agreements are also available in many areas.

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals