Front Panel Operation
Introduction to Offset, Scale, and Linearity Errors 4
4-41
4-42. Reading the UUT Error: Resistance Output
Proceed as follows to read the error of the UUT in the resistance function:
1.
Set the desired resistance output as described under “Setting the Output.”
2.
Use the output adjustment controls as required to achieve a reading on the Control
Display (illustrated below) equal to the
reading on the UUT. To adjust a higher-order
digit, press
<. As you approach the reference value, work progressively back
towards the LSD (least-significant digit) on the calibrator’s Control Display as
necessary by pressing
>. You only need to go one digit past the LSD of the UUT;
digits below that are beyond the
resolution of the UUT.