EasyManua.ls Logo

Fluke 5700A Series II - Cables; Guarding; Sensing

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5700A/5720A Series II
Operators Manual
4-52
4-55. Cables
According to the text under “Connecting the Calibrator to a UUT,” the 5440A-7002
(banana plugs) or 5440A-7003 (spade lugs) Low-Thermal Cables are recommended for:
All DC voltage
All resistance
AC voltage <10 kHz
AC current 2 A 0 kHz
DC current 2A
Since all of the 70 Series tests include these functions, you should use the 5440A-7002 or
5440A-7003 Low-Thermal Cables, if possible. However, the thermal emf errors that the
Low-Thermal cables are designed to reduce are not significant when calibrating a 3½
digit meter. Therefore, if you have the recommended Low-Thermal Cables, use them.
Otherwise, ordinary insulated test leads will suffice.
4-56. Guarding
Because the 70 Series Multimeters are battery operated, their inputs have no connection
to earth ground. Therefore, setting the calibrator to internal guard is appropriate.
4-57. Sensing
External sensing is required only when the voltage drop in the cables is significant. The
following data determines the significance of the voltage drop in the cables:
The input impedance of the 70 Series Multimeters is 10 M.
The total resistance of the Model 5440A-7002 Low-Thermal Cables is 30 m.
The basic dc voltage accuracy of the most accurate of the 70 Series, the Model 77, is
0.3%.
The percentage of error caused by voltage drop in the cables is:
0.3 x 100/10,000 = 0.003%, or 1/100th of the accuracy of the Model 77.
This voltage drop is not significant, so a two-wire connection (Internal Sense) is
appropriate.

Table of Contents

Other manuals for Fluke 5700A Series II

Related product manuals