EasyManua.ls Logo

Fluke 5700A Series II - Calibration Check

Fluke 5700A Series II
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5700A/5720A Series II
Operators Manual
5-64
5-51. Calibration Check
1Ø ! THIS IS A PROGRAM FOR REMOTE CALIBRATION CHECK
2Ø !
3Ø REMOTE
4Ø PRINT @4, “*CLS; *SRE 8” ! ENABLE EAV TO TRIGGER SRQ
5Ø ON SRQ GOTO 7Ø ! SRQ HANDLER
6Ø GOTO 15Ø ! START MAIN PROGRAM
7Ø PRINT @4, “FAULT?” \ INPUT @4, A% ! SRQ HANDLER, GET FAULT CODE
8Ø IF A% = Ø GOTO 13Ø ! IF NO FAULT EXIT
9Ø PRINT @4, “EXPLAIN? “;A% \ INPUT @4, A$ ! GET EXPLANATION
1ØØ PRINT
11Ø PRINT “FAULT”;A%;”OCCURRED: “;A$ ! DISPLAY EXPLANATION
12ØG OTO 7Ø ! GET NEXT FAULT
13Ø A% = SPL(4%) ! SERIAL POLL TO CLEAR SP BYTE
14Ø RESUME 3ØØ ! END OF SRQ HANDLER
15Ø PRINT “Ambient temperature in Celsius”; ! MAIN PROGRAM BEGINS
16Ø INPUT T ! GET TEMPERATURE FROM OPERATOR
17Ø PRINT @4, “CAL_TEMP”;T ! SET CAL TEMPERATURE
18Ø PRINT “Doing calibration check. Please wait . . .”
19Ø PRINT @4, “CAL_CHK;*OPC” \ GOSUB 27Ø
2ØØ PRINT “Calibration check finished. Print a report? [y, n]”
21Ø INPUT A$ \ IF LEFT(A$,1)=“N” OR LEFT(A$,1)=“n” GOTO 3ØØ
22Ø PRINT “Set up the peripheral device.” \ GOSUB 25Ø
23Ø PRINT “Printing ...” \ PRINT @4, “CAL_PR CHECK;*OPC” \ GOSUB 27Ø \ GOTO
3ØØ
25Ø PRINT “Press RETURN to continue” ! PAUSE FOR OPERATOR SUBROUTINE
26Ø INPUT A$ \ RETURN ! “
27Ø PRINT @4, “*ESR?” \ INPUT @4,A% ! WAIT FOR OPC SUBROUTINE
28Ø IF A% AND 1% THEN RETURN ! “, RETURN IF OPC BIT IS 1
29Ø GOTO 27Ø ! “, STAY IN LOOP IF OPC IS Ø
3ØØ PRINT “All finished!” \END

Table of Contents