Front Panel Operation
Setting the Output 4
4-29
8. Press up to five numeric keys to select a frequency (followed by K if necessary).
The Control Display now shows the amplitude and frequency of your entry. If you
made an entry error, press C to clear the display, then reenter the value. The
following illustration of the Control Display assumes an entry of 2.5 kHz:
PREV
MENU
9. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display. No current is available at the output terminals, however,
until you press O.
10. Press O to activate the calibrator’s output. The UUT will now respond to the
applied current.
Three softkey labels appear on the Control Display in the ac current function: “Currnt
Output”, “Phase Ctrls Menu”, and “Setup Menus” as shown below:
PREV
MENU
The “Currnt Output” softkey selects one of three locations for current output: the
OUTPUT binding post (NORMAL, which is the OUTPUT HI binding post, also the
default setting); AUX, which is the AUX CURRENT OUTPUT binding post, or
5725A, which is the 5725A Amplifier binding posts. (The 5725A must be turned on,
but not necessarily activated.)
Note
If AUX is selected for a calibrator configured for rear output, error
message: “Compliance voltage exceeded” will be displayed. For rear
output, use the NORMAL setting.