Remote Operation
Commands 5
5-19
Table 5-2. Commands (continued)
CAL_CONF
Description Sequential command.
Sets the calibration specifications to a confidence level of either 99% or 95%.
Parameters:
CONF95 or CONF99: Sets 95% or 99% specifications.
CAL_CONF?
Description Sequential command.
Returns the current calibration confidence level.
Parameters:
None
Response CONF95 (indicates 95% specifications)a or CONF99 (indicates 99% specifications)
CAL_CONST?
Description Sequential command.
Returns the value of a particular calibration constant from a particular group of constants.
Parameters:
1. CAL (From active calibration constants)
CHECK (From calibration check constants)
PREV (From previous calibration constants)
2. Symbolic name of the calibration constant desired (see Appendix D).
Response (Float) The value of the constant
Example: CAL_CONST? CHECK, KV6
Could return: 6.5000010E+00
CAL_DATE?
Description Sequential command.
Returns the date of the most recent calibration of the specified type.
Parameter
B5725 (Last 5725A Amplifier calibration)
CAL (Last calibration of the calibrator)
WBFLAT (Last wideband flatness calibration)
WBGAIN (Last wideband gain calibration)
ZERO (Last dc V zero calibration)
Response (Integer) Date as MMDDYY, DDMMYY, or YYMMDD depending on the setting of
DATEFMT.
CAL_DAYS?
Description Sequential command.
Returns the number of days elapsed since the last calibration activity of the specified type.
Parameter
B5725 (Last 5725A Amplifier calibration)
CAL (Last calibration of the calibrator)
WBFLAT (Last wideband flatness calibration)
WBGAIN (Last wideband gain calibration)
ZERO (Last dc V zero calibration)
Response (Integer) Number of elapsed days