Remote Operation
Commands 5
5-21
Table 5-2. Commands (continued)
CAL_RNG
Description Overlapped command, ignored if not in remote.
Starts a calibration of a range. This command causes the calibrator to source the calibration
magnitude specified by the second parameter for the range specified by the first parameter.
To calibrate a range, the controller must send commands in the same sequence as in the
example.
Parameters:
1. The range identifier of the range to calibrate from Table 5-4, or the keyword NULL or
STORE.
2. (Only if parameter 1 is the range mnemonic) The calibration magnitude for the calibrator
to source. Must be 45% to 95% of the range specified.
3. (Optional; only if parameter 1 is the range mnemonic) The frequency of the calibrator
output with optional multiplier and units.
Example:
CAL_TEMP 23.5
CAL_RNG DC2_2V, 2V ; OPER ; *WAI
INCR (enough to null the calibrator)
CAL_RNG NULL
CAL_RNG STORE
CAL_RPT?
Description Sequential command.
Returns a report for a specified calibration activity (see Chapter 6 for format details).
Parameter CAL (Output change report for calibration)
CHECK (Output change report for cal check)
RAW (A list of all calibration constants)
Response (String)
"<EOL>
<formatted report>
"
Example: 10 REM THIS 1722A PROGRAM SAVES FORMATTED REPORT IN A FILE
20 CLOSE 1 \ OPEN "MF0:CALRPT" AS NEW FILE #1
30 PRINT @4,'CAL_RPT? CAL'
40 INPUTLINE @4,A$ ! THROW AWAY BEGINNING QUOTE
50 ! DO {
60 INPUTLINE @4,A$
70 PRINT #1,A$
80 IF LEFT(A$,1%) != '"' GOTO 60
90 !} WHILE(NOT END OF FILE)