EasyManua.ls Logo

Fluke 5700A Series II - Page 173

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
Commands 5
5-23
Table 5-2. Commands (continued)
CAL_SLST?
Description Sequential command.
Returns a group of calibration constant shifts due to a calibration activity.
Parameter CAL (All output changes due to calibration)
CHECK (All output changes due to calibration check)
Response (String)
"<EOL>
<range name>,<# points><EOL>
<mag 1>,<freq 1>,<offset 1>,<ashift 1>,<rshift 1><sshift 1>,<spec 1><EOL>
. . .
<mag n>,<freq n>,<offset n>,<ashift n>,<rshift 1>,<sshift n>,<spec n><EOL>
"
Where:
<range name> = Range identifier from Table 5-4
<# points> = (Integer) Number of points for the range
<mag n> = (Float) Magnitude for point n in range units
<freq n> = (Float) Frequency for point n in Hz
<offset n> = (Float) Zero shift for point n in range units
<ashift n> = (Float) Absolute shift for point n in range units
<rshift n> = (Float) Relative shift for point n (ppm)
<sshift n> = (Float) % of spec shift for point n
<spec n> = calibrator specification for point n in ppm
Range units are the appropriate units for the range, e.g., for all DC V ranges, range units
are V.
Example:
CAL_SLST? CAL could return:
49
DC220MV,2
2.20E-1,0.00E+00,1.76E-07,1.97E-07,8.98E-01,7.10E+00,1.26E+01
-2.20E-1,0.00E+00,1.58E-07,1.38E-07,6.26E-01,4.95E+00,1.26E+01
DC2_2V,2
(cont)
"
CAL_STORE
Description Sequential command, ignored if not in remote.
Stores all new calibration constants in nonvolatile memory. Use this command to save the
results of a CAL_ADJ command after you are completely finished with calibration via remote
control. The rear panel CALIBRATION switch must be in the ENABLE position or an
execution fault results.
Parameter None

Table of Contents

Other manuals for Fluke 5700A Series II

Related product manuals