Using the RS-232C Serial Interface
Printing Calibration Reports 6
6-9
-----------------------------------------------------------------------------
FLUKE CORPORATION
A S/N 1824760B 5720A OUTPUT CALIBRATIONC
PRINTED ON 12/14/95 AT 14:04:24 5720A
-----------------------------------------------------------------------------
-----------------------------------------------------------------------------
MODULES PRESENT
-----------------------------------------------------------------------------
Software Revision 0.2+A+*
D
Switching Matrix
DC Volt Module
AC Volt Module
220V Module
E
1100V/2A Module
Current Module
Ohms Module
Wideband Module
Hires Osc Module
-----------------------------------------------------------------------------
ARTIFACT CALIBRATION DATES AND TEMPERATURES
-----------------------------------------------------------------------------
Specification Interval: 90day Confidence 99%
MAIN OUTPUT: (Most recent) 12/06/88, 22.9 Degrees C
F
(Previous) 9/07/88, 23.2 Degrees C
(Days since last cal) 43
ZERO: (Most recent) 1/05/89
(Days since last cal) 13
WIDEBAND GAIN: (Most recent) 12/06/88, 23.0 Degrees C
G
(Previous) 9/07/88, 23.3 Degrees C
(Days since last cal) 43
WIDEBAND FLAT: (Most recent) 12/06/88, 22.9 Degrees C
H
(Previous) 9/07/88, 23.3 Degrees C
(Days since last cal) 43
-----------------------------------------------------------------------------
STANDARDS USED
-----------------------------------------------------------------------------
STANDARD NUMBER CERT. DATE
VOLTAGE ______________ ______________ 10.000017 V
10 kOhm ______________ ______________ 10.000124 kOhm
I
1 Ohm ______________ ______________ 1.0000237 Ohm
OTHER ______________ ______________ _______________
______________ ______________ _______________
-----------------------------------------------------------------------------
EXTERNALLY CALIBRATED INTERNAL REFERENCES
-----------------------------------------------------------------------------
REFERENCE Most Recent Previous Shift (ppm)
6.5 V DC Reference 6.548268 6.548268 0.00
13 V DC Reference 13.082727 13.082725 0.20
10 kOhm Reference 1.002937 1.002938 -0.11
J
1.0 Ohm Resistor 1.0000967 1.0000980 -1.32
1.9 Ohm Resistor 1.8992432 1.8992446 -0.72
K L M
Figure 6-2. Calibration Shift Results Printout

Table of Contents

Other manuals for Fluke 5700A Series II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5700A Series II and is the answer not in the manual?

Fluke 5700A Series II Specifications

General IconGeneral
TypeMultifunction Calibrator
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 70 °C
Voltage Range0 to 1100 V
DC Voltage Range10 μV to 1100 V
Frequency Range1 MHz
Humiditynon-condensing
Power Requirements100 V to 240 V

Related product manuals