EasyManua.ls Logo

Fluke 5700A Series II - Page 257

Fluke 5700A Series II
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator Maintenance
Range Calibration 7
7-15
3. Press the “Range Adjust” softkey to bring up the next menu shown below:
PREV
MENU
4. Press the “DC V” softkey to bring up the next menu shown below:
PREV
MENU
5. Press the “NEXT Menu” softkey, which scrolls through all the available dc voltage
ranges, until 220V appears. Then press that selection’s softkey to open a display
similar to the following:
PREV
MENU
6. Connect the 732B, 845A, and 732B in a 10:1 configuration, as shown in Figure 7-5.
7. Multiply the 732B’s value by 10, and enter this new value. (This value is the output
of the 752A, to which you will null the calibrator’s output.) Then press E to
bring up the following display.
PREV
MENU

Table of Contents

Other manuals for Fluke 5700A Series II

Related product manuals